Transcatheter closure of perimembranous ventricular septal defects with Amplatzer occluder assessed by real-time three-dimensional echocardiography
Unité de Cardiologie Pédiatrique, Hôpital des Enfants, 330 avenue de Grande-Bretagne, B-P 3119, 31026 Toulouse Cedex 3, France
Received 21 November 2005; received in revised form 14 February 2006; accepted after revision 28 February 2006.
* Corresponding author. Tel.: +33 5 34 55 87 34; fax: +33 5 34 55 86 63. acar.p{at}chu-toulouse.fr
| Abstract |
|---|
|
|
|---|
We report our initial experience in two patients using real-time three-dimensional echocardiography to assess perimembranous ventricular septal defect and device morphology and their relation with contiguous cardiac structure. Defect size and rims as well as device position and profile were displayed from the three-dimensional "en face" views. We think that real-time three-dimensional echocardiography could be a complementary approach to angiography and transesophageal echocardiography in performing transcatheter closure of perimembranous ventricular septal defect.
Keywords: Three-dimensional echocardiography; Transcatheter closure; Perimembranous ventricular septal defect; Amplatzer occluder
| Introduction |
|---|
|
|
|---|
The isolated perimembranous ventricular septal defect (PM VSD) is one of the most common congenital heart malformations. Treatment has been indicated in the presence of significant left-to-right shunt, resulting in left ventricular overload. Surgical closure has been performed with a very low mortality rate, although complications, such as residual leaks, atrioventricular block, postpericardiotomy syndrome and arrhythmias have been well described. Transcatheter closure PM VSD has been attempted as an alternative approach to surgery.1–4 The specific device has to be positioned very closely to the aortic and tricuspid valve. The device closure procedure is performed using in a combination of fluoroscopic and transesophageal echocardiography (TEE). Since two-dimensional (2D) echocardiography approaches the ventricular septum from multiple orthogonal planes, it requires a mental three-dimensional (3D) construct to comprehend the relation between the defect and the surrounding structures. Three-dimensional echocardiography (3D echo) provides unique "en face" views of the atrial and ventricular septum.5,6 We report our initial experience in two patients using 3D echo to assess PM VSD and device morphology, and their relation with contiguous cardiac structure.
| Case report |
|---|
|
|
|---|
Population
Two consecutive patients were referred for device closure of their PM VSD. Age and weight were respectively 2years and 11kg (patient 1), 3.5years and 12kg (patient 2). Each patient underwent 2D transthoracic echocardiography. Size of PM VSD, QP/QS ratio and left ventricular diastolic diameter were respectively 7mm, 2.2 and 38mm (patient 1), 9mm, 2.4 and 40mm (patient 2).
3D echocardiography
Live 3D echo was performed the day before and after the transcatheter procedure using a Sonos 7500 system (Philips, Andover, MA). The 2–4MHz matrix probe has a beam-former connecting 3000 piezoelectric elements allowing 3D images. The system acquired 3D data using a 40°x20° pyramidal imaging volume (live 3D). Cardiac cycles were also acquired using the "full-volume" mode that consists of the acquisition of a larger single pyramid of data (120°x60°) recorded during four consecutive cardiac cycles. Navigation by cropping inside the volume allowed surface rendering of the intracardiac structures. Measurement of VSD diameter was obtained using the Q-Lab system (Philips).
Transcatheter procedure
The procedure was performed under general anesthesia and TEE. Angiography in single plane (50° LAO/15° cranial) was performed to define the location and size of the PM VSD. The VSD was crossed from the left ventricle and a guidewire advanced into the pulmonary artery. A floppy exchange guidewire was advanced through the arterial catheter, snared and withdrawn from the femoral vein, establishing through-and-through access. A 9-F dilator and delivery Mullins sheath was advanced from the femoral vein across the ventricular septal defect into the ascending aorta.
Device selection
The PM VSD size was measured by TEE, 3D echo and angiography. Maximal diameter of the defect was measured on the left ventricular side in diastole as required.7 The data in two patients are reported in Table 1. The Amplatzer membranous VSD occluder (AGA, MN, USA) designed specifically for PM VSD has a diameter ranging from 4 to 18mm with 2mm increments. The device was selected to be 1–2mm larger than the VSD size estimated by both TEE and angiography without taking account of 3D data. The device size was respectively 10 in patient 1 and 14 in patient 2.
|
Device delivery
In patient 2, the sheath tip was withdrawn to the left ventricular outflow tract and the guidewire was advanced to the LV apex with the arterial catheter. The sheath and dilator were advanced to the LV apex. In patient 1, the sheath could not be positioned in the LV apex; therefore, it was left in the ascending aorta and the device was deployed just adjacent to the aortic valve. The left ventricular disc was deployed under fluoroscopic control. Inferior direction of the platinum disc marker was confirmed. Under TEE guidance, the left ventricular disc was withdrawn to the septum. Good position was confirmed by left ventriculography. Absence of aortic insufficiency was confirmed by color flow Doppler. Keeping a steady traction on the delivery rod, the right side of the device was then deployed on the right ventricular side of the defect. Good position was confirmed by TEE and angiography. Tricuspid, mitral, and aortic insufficiency were sought on color flow Doppler. Any arrhythmia or conduction defects were noted. The device was released by unscrewing the microscrew. A left ventriculogram and aortogram were performed and showed a small shunt through the device.
Follow-up
Device position was assessed by 2D and 3D echocardiography the day after the procedure. Any residual shunt and aortic regurgitation was observed. In patient 1, the profile of the device was flat. In patient 2, probably due to a slight oversizing, the device had a mild "mushroom" profile without left ventricular outflow Doppler gradient. The patients recovered in an appropriate setting and were discharged home the following day. Patients received 100mg aspirin daily for 6months. During a follow-up of 3months after device release, any arrhythmia or conduction defects were noted.
| Discussion |
|---|
|
|
|---|
Real-time 3D echocardiography is a complementary tool to assess anatomy of PM VSD. Size of the defect and relation with contiguous cardiac structure are two crucial parameters to evaluate prior transcatheter closure.1–4 Device selection is based on VSD diameter measurements. Conventional methods such as angiography and 2D echocardiography do not integrate the complex and three-dimensional structure of the ventricular septum, which could explain the differences in VSD measurement between these methods. We found in the two patients that PM VSD diameter by 3D echo was 2mm superior to diameter by TEE. Such a finding is explained by the asymmetric shape of the defects (Figs. 1 and 2
|
|
The 3D "en face" views allowed better comprehension of the VSD and its relation with adjacent structures (Figs. 3 and 4
|
|
The geometric profile of the device has to be specifically designed for PM VSD closure. A variety of occluding devices, such as the Rashkind double umbrella, the buttoned device, or coils, were unsuccessfully used to close PM VSD. The Amplatzer membranous VSD occluder has a specific profile consisting of two parallel discs with minimal subaortic rim. The left disc is eccentric, 0.5mm larger than the waist at its superior rim and 5.5mm at its inferior rim. The right disc is centric, being 4mm round the waist. Authors focused on the volume of the Amplatzer device in the atrial cavity and its potential obstruction. We presently report the two first PM VSD occluders assessed by 3D echo. The 3D views illustrated the geometric profile of the Amplatzer device in both patients (Figs. 5–8
|
|
|
|
McKendrick et al. reported the use of real-time 3D echo to guide device closure of atrial septal defect.10 They concluded that this method was a feasible, safe and effective alternative to the standard practice of TEE. Higher spatial resolution and frequency probe are required to improve 3D image quality. The freeze 3D images dampened partially the strength of 3D echo study, which needs to be displayed in dynamic format. However, we think that real-time 3D echo could be a complementary approach in planning for and performing transcatheter closure of PM VSD. Further studies with a larger number of patients are required to assess the place of each imaging modality.
| References |
|---|
|
|
|---|
- Hijazi Z.M., Hakim F., Haweleh A.A., Madani A., Tarawna W., Hiari A., et al. Catheter closure of perimembranous ventricular septal defects using the new Amplatzer membranous VSD occluder: initial clinical experience. Catheter Cardiovasc Interv (2002) 56:508–515.[CrossRef][Web of Science][Medline]
- Pedra C.A., Pedra S.R., Esteves C.A., Pontes S.C. Jr., Braga S.L., Arrieta S.R., et al. Percutaneous closure of perimembranous ventricular septal defects with the Amplatzer device: technical and morphological considerations. Catheter Cardiovasc Interv (2004) 61:403–410.[CrossRef][Web of Science][Medline]
- Fu Y.C., Bass J., Amin Z., Radtke W., Cheatham J.P., Hellenbrand W.E., et al. Transcatheter closure of perimembranous ventricular septal defects using the new amplatzer membranous VSD occluder. J Am Coll Cardiol (2006) 47:319–325.
[Abstract/Free Full Text] - Carminati M., Butera G., Chessa M., Drago M., Negura D., Piazza L. Transcatheter closure of congenital ventricular septal defects with amplatzer septal occluders. Am J Cardiol (2005) 96:52–58.[Web of Science][Medline]
- Acar P., Saliba Z., Bonhoeffer P., Aggoun Y., Bonnet D., Sidi D., et al. Influence of atrial septal defect anatomy in patient selection and assessment of closure by the Cardioseal device: a three-dimensional transesophageal echocardiography study. Eur Heart J (2000) 21:573–581.
[Abstract/Free Full Text] - Acar P., Abdel-Massih T., Douste-Blazy M.Y., Dulac Y., Bonhoeffer P., Sidi D. Assessment of muscular ventricular septal defect closure by transcatheter or surgical approach: a three-dimensional echocardiographic study. Eur J Echocardiogr (2002) 3:185–191.[CrossRef][Medline]
- Cheng T.O., Xie M.X., Wang X.F., Wang Y., Lu Q. Real-time 3-dimensional echocardiography in assessing atrial and ventricular septal defects: an echocardiographic-surgical correlative study. Am Heart J (2004) 148:1091–1095.[CrossRef][Web of Science][Medline]
- Ho S.Y., McCarthy K.P., Rigby M.L. Morphology of perimembranous ventricular septal defects: implications for transcatheter device closure. J Interv Cardiol (2004) 17:99–108.[CrossRef][Medline]
- Scheurer M., Bandisode V., Ruff P., Atz A., Shirali G. Early experience with real-time three-dimensional echocardiographic guidance of right ventricular biopsy in children. Echocardiography (2006) 23:45–52.[CrossRef][Web of Science][Medline]
- McKendrick R., Owada C.Y. Real-time 3D echocardiography-guided transcatheter device closure of atrial septal defects. Catheter Cardiovasc Interv (2005) 65:442–446.[CrossRef][Web of Science][Medline]
This article has been cited by other articles:
![]() |
K. Ratnayaka, V. K. Raman, A. Z. Faranesh, M. Sonmez, J.-H. Kim, L. F. Gutierrez, C. Ozturk, E. R. McVeigh, M. C. Slack, and R. J. Lederman Antegrade Percutaneous Closure of Membranous Ventricular Septal Defect Using X-Ray Fused With Magnetic Resonance Imaging J. Am. Coll. Cardiol. Intv., March 1, 2009; 2(3): 224 - 230. [Abstract] [Full Text] [PDF] |
||||
![]() |