Late cardiac tamponade after percutaneous closure of a patent foramen ovale
aDivision of Cardiology, University Hospital, Geneva, Switzerland
bDivision of Radiology, University Hospital, Geneva, Switzerland
cDivision of Cardiovascular Surgery, University Hospital, Geneva, Switzerland
Received 30 March 2005; received in revised form 25 July 2005; accepted after revision 3 August 2005.
pedro.trigo-trindade{at}hcuge.ch
* Corresponding author. Division of Cardiology, Geneva University Hospital, 24 rue Micheli-du-Crest, 1211 Geneva 14, Switzerland. Tel.: +41 22 372 71 95; fax: +41 22 372 72 29.
| Abstract |
|---|
|
|
|---|
We report the case of a 35-year-old man who had a transient ischemic cerebral attack and then underwent a percutaneous closure of the patent foramen ovale (PFO) with a Cardiastar device. One year later, the patient developed a cardiac tamponade due to an important hemorrhagic pericardial effusion. Transoesophageal echocardiography showed that one of the struts had impinged on the aortic root in the region adjacent to the transverse pericardial sinus. Therefore, we speculated that the strut had passed through the aortic wall by slow erosion, leading to the pericardial effusion. Cardiac CT and subsequent surgery confirmed the perforation of the left atrial roof and the aortic root by two struts of the device. This is the first reported case of late cardiac tamponade and underscores the importance of long-term follow-up after PFO closure device implantation.
Keywords: Atrial septal abnormality; Percutaneous PFO closure; Cryptogenic stroke; Pericardial effusion; Late complication; Intracardiac device
| Introduction |
|---|
|
|
|---|
A 35-year-old man with no previous medical problems was admitted to our hospital in February 2003 because of a transient ischemic cerebral attack. A small left frontal ischemic lesion was detected with a brain CT-scan. Furthermore, we found a patent foramen ovale (PFO), which had a significant right-to-left shunt that was associated with an atrial septal aneurysm, by transoesophageal echocardiography (TEE) (Fig. 1A and B). The distances from the PFO to the superior vena cava and the aortic root were 12mm and 7mm, respectively (Fig. 1C and D). Following neurological recovery, we used a Cardiastar device (3/30mm, third generation) to perform a percutaneous closure of the PFO. Besides fluoroscopy, other imaging techniques were not performed. The patient was discharged after the uneventful percutaneous PFO closure and he returned six months later for a follow-up TEE. This exam did not indicate any pathological mobility or perforation of the intracardiac device, but did show that it was in contact with the aortic root (Fig. 2A). Furthermore, a very small residual right-to-left shunt during the Valsalva manoeuvre was also detected (data not shown).
|
|
The patient returned to our hospital one year later due to a sudden onset of left thoracic pain while at rest, which was followed by dizziness and syncope. On examination, his jugular veins were raised and he developed cardiogenic shock that required mechanical ventilation and Adrenaline i.v. Because an important pericardial effusion with compression on the right cardiac chambers was detected by a transthoracic echocardiography (Fig. 2B), we performed an emergency pericardial puncture and removed 250ml of hemorrhagic fluid. An aortic dissection was ruled out, first by a thoracic CT-scan (data not shown) and then by a TEE. Importantly, the TEE showed no pathological mobility of the Cardiastar device, but it indicated that one of the struts had impinged on the aortic root, in a region adjacent to the transverse pericardial sinus (Fig. 2C). Consequently, we speculated that the left atrium and the proximal aorta had been perforated, which resulted in the pericardial effusion. In fact, in comparison with the first TEE (Fig. 2A), the struts had spread open and pushed against the aortic root (Fig. 2C). By cardiac CT-scan, we confirmed that the Cardiastar's cranial strut had passed through the aortic wall (Fig. 3). During removal of the Cardiastar device by open-heart surgery, we detected two nitinol struts from the Cardiastar device, which had perforated the roof of the left atrium (Fig. 4) and one of these two had perforated into the non-coronary sinus of Valsalva of the aortic root. The patient recovered uneventfully from this surgery and was asymptomatic one year later.
|
|
| Discussion |
|---|
|
|
|---|
Based on autopsy reports,1 the prevalence of PFO is 17–35% in the general population. While numerous studies have suggested an association between PFO and cryptogenic stroke, there are limited explanations on how atrial septal abnormalities are implicated in cryptogenic stroke.2 In spite of this, percutaneous PFO closure in patients with a paradoxical embolism has emerged as an alternative treatment strategy to lifelong anticoagulation, antiplatelet agents or surgical PFO closure.3 Several peri-procedural and short-term complications have been recognized in PFO device implantation, but their rate is low.4 Pericardial effusion after percutaneous PFO closure has been reported as a rare early complication. This complication was described with different devices.5,6 In this case, we suspected that slow erosion of the left atrial roof and the aortic wall were responsible for a hemorrhagic pericardial effusion. There was no shunt detected between the aorta and the left atrium because the responsible strut remained in place until surgical removal. We believe that the described complication is due to both the device and its position. In fact, star-like devices have been associated with a higher rate of complications, including early tamponade, when compared to stent-like devices.5 Furthermore, the PFO in our patient was close to the aortic root, which makes the implantation of a device more difficult, and the use of a star-like device possibly more dangerous.
In conclusion, this case illustrates that TEE or intracardiac echo imaging during device implantation is advisable to evaluate the position of the umbrella with respect to the aortic root. Furthermore, this report shows that pericardial effusion can occur even one year after percutaneous PFO closure, which highlights the importance of long-term follow-up of patients after device implantation.
| References |
|---|
|
|
|---|
- Hagen P.T., Scholz D.G., Edwards W.D. Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc (1984) 59:17–20.[Web of Science][Medline]
- Messe S.R., Silverman I.E., Kizer J.R., Homma S., Zahn C., Gronseth G., et al. Practice parameter: recurrent stroke with patent foramen ovale and atrial septal aneurysm: report of the Quality Standards Subcommittee of the American Academy of Neurology. Neurology (2004) 62:1042–1050.
[Abstract/Free Full Text] - Meier B., Lock J.E. Contemporary management of patent foramen ovale. Circulation (2003) 107:5–9.
[Free Full Text] - Braun M., Gliech V., Boscheri A., Schoen S., Gahn G., Reichmann H., et al. Transcatheter closure of patent foramen ovale (PFO) in patients with paradoxical embolism. Periprocedural safety and mid-term follow-up results of three different device occluder systems. Eur Heart J (2004) 25:424–430.
[Abstract/Free Full Text] - Schwerzmann M., Windecker S., Wahl A., Mehta H., Nedeltchev K., Mattle H., et al. Percutaneous closure of patent foramen ovale: impact of device design on safety and efficacy. Heart (2004) 90:186–190.
[Abstract/Free Full Text] - Trepels T., Zeplin H., Sievert H., Billinger K., Krumsdorf U., Zadan E., et al. Cardiac perforation following transcatheter PFO closure. Catheter Cardiovasc Interv (2003) 58:111–113.[CrossRef][Web of Science][Medline]
Related articles in Eur J Echocardiogr:
- Hazards of percutaneous PFO closure
- Markus Schwerzmann and Omid Salehian
Eur J Echocardiogr 2005 6: 393-395.[Extract] [FREE Full Text]
This article has been cited by other articles:
![]() |
K. Yared, A. L. Baggish, J. Solis, R. Durst, J. J. Passeri, I. F. Palacios, and M. H. Picard Echocardiographic Assessment of Percutaneous Patent Foramen Ovale and Atrial Septal Defect Closure Complications Circ Cardiovasc Imaging, March 1, 2009; 2(2): 141 - 149. [Full Text] [PDF] |
||||
![]() |
D Kenny, M Turner, and R Martin When to close a patent foramen ovale Arch. Dis. Child., March 1, 2008; 93(3): 255 - 259. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cecconi, A. Quarti, F. Bianchini, S. Bucari, C. Costantini, A. Giovagnoni, and G. P. Perna Late Cardiac Perforation After Transcatheter Closure of Patent Foramen Ovale Ann. Thorac. Surg., June 1, 2006; 81(6): e29 - e30. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||