High Frame Rate Myocardial Integrated Backscatter. Does this Change our Understanding of this Acoustic Parameter?
,2
1Medical Image Computing, Department of Electrical Engineering, K.U. Leuven, Belgium
2Department of Cardiology, K.U. Leuven, Belgium
3Laboratorium voor akoestiek en thermische fysica, Department of Physics, K.U. Leuven, Belgium
4Department of Cardiology, Klinikum Innenstadt, University of Munchen, Germany
| Abstract |
|---|
Aims: Integrated backscatter (IB) and its cyclic variation (CV) derived from radio-frequency (RF) data have been used as parameters to attempt myocardial tissue characterization. Prior imaging systems used to measure IB and its CV typically acquired data at frame rates of 20–30 Hz and at a resolution of 6–8 bits. If changes in IB levels are in part related to specific short-lived events, occurring within the cardiac cycle, this frame rate and resolution could have been too low to resolve adequately what might be a more complex data set.
Methods and Results: To investigate this possibility, we acquired real time two-dimensional (2D) myocardial IQ data (the in-phase quadrature sampled RF data) at high frame rate (>100 Hz), high dynamic resolution (theoretical 19-bit) and a sector angle of 20°. Several consecutive heart cycles of myocardial data were acquired from individual cardiac walls in five closed chest dogs and 10 healthy, young volunteers at normal heart rates. On the reconstructed RF data regions of interest were indicated, and IB and its CV were calculated.
The extracted high frame rate curves showed that the CV of IB is not a smooth sinusoidal-like curve, but is made up of multiple reproducible peaks and troughs with local minima and maxima which are temporally related to active or passive mechanical events, i.e. systolic contraction, early ventricular relaxation and ventricular filling due to atrial contraction.
Conclusions: This study shows that increasing the rate of real-time RF data acquisition results in a more complex, reproducible IB curve. The resolved maxima and minima in IB levels are related to specific phases of the myocardial contraction. Furthermore, spectral analysis showed that IB curves acquired at normal heart rates contain information up to 40 Hz. Hence, cardiac imaging data sets used to analyse regional myocardial function obtained at frequencies lower than 80 frames per second can contain aliased information.
Keywords: Cyclic variation; Integrated backscatter; Temporal resolution; Tissue characterization
* Correspondence address: Jan D'hooge, UZ Gasthuisberg, Department of Cardiology, Herestraat 49, B-3000 Leuven, Belgium.
G. R. Sutherland is a Visiting Professor of Cardiology at the K.U. Leuven.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
A. Micari, M. Pascotto, A. R. Jayaweera, J. Sklenar, N. C. Goodman, and S. Kaul Cyclic variation in ultrasonic myocardial integrated backscatter is due to phasic changes in the number of patent myocardial microvessels. J. Ultrasound Med., August 1, 2006; 25(8): 1009 - 1019. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.E Hancock, J.C Cooke, and M.J Monaghan Effect of harmonic imaging on the measurement of ultrasonic integrated backscatter and its interpretation in patients following myocardial infarction Eur J Echocardiogr, June 1, 2004; 5(3): 189 - 195. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Weidemann, F. Jamal, G. R. Sutherland, P. Claus, M. Kowalski, L. Hatle, I. De Scheerder, B. Bijnens, and F. E. Rademakers Myocardial function defined by strain rate and strain during alterations in inotropic states and heart rate Am J Physiol Heart Circ Physiol, August 1, 2002; 283(2): H792 - H799. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Hatle and G. R. Sutherland Regional myocardial function--a new approach. Eur. Heart J., August 1, 2000; 21(16): 1337 - 1357. [PDF] |
||||